670 Electrolysis Workstation
670 Electrolysis Workstation
The 670 Electrolysis Workstation is a comprehensive, customizable platform capable of accommodating both liquid and gas on one or both electrodes. It features a multi-range, high-current (±20 Amp) potentiostat, ensuring high accuracy across a wide range of reaction rates and cell sizes. The system includes the powerful and user-friendly FlowCell® software for system operation and experimentation. With numerous accessory options available, you can tailor this system to suit your specific research needs. It is particularly well-suited for Electrolysis and Desalination by Capacitive Deionization studies, as well as applications such as Power-to-Fuels & Power-to-Chemicals, CO2 Reduction, Low Temperature CO & Syngas Production, Electrochemical Ammonia Synthesis, and Organic Electrosynthesis.
לפרטים נוספים והצעת מחיר, מלא את הפרטים למטה
מידע נוסף
Scribner 670 Electrolysis Workstation Data Sheet
- Multi-range potentiostat – offering ±70mA/700mA/7A/20A, ±5V capabilities
- Integrated 881 FRA for EIS & HFR analysis
- Heated & Temperature Controlled Borosilicate Bottles – available in sizes of 100, 250, 500 & 1,000 mL
- Teflon Stir Bar provided in both positive and negative tanks
- Custom Flow Bench includes:
- Software Controlled Peristaltic Pumps
- Heated Pressure-rated Duran Bottles
- Designed for easy removal, fill, purge, rinse, and drain
- N2 Reservoir Backfill & System Purge with Check Valves
- All Non-Metal Wetted Parts
- Taps For Sampling
- Options for Dry or Humidified Gas
- Gas System for seamless switching between gases
- Includes custom FlowCell®, BView®, and ZView® software
- Offers REF inputs for ½-cell measurements, including EIS and HFR
- Manual Back Pressure System to 1.5 bar gauge with pressure-rated bottles. Pressure monitoring integrated with FlowCell® Software with addition of 892 Data Expansion Unit and pressure transducers.
- Scribner Flow Cell Test fixture
- 5 or 25 cm2 cells with graphite and/or Pt-Ti flow fields
- 892 Data Expansion Unit
Potentiostat: | |
Current Ranges: | ±20 A, 7 A, 700 mA, 70 mA full scale (FS) |
Current Resolution: | 1.4 mA (20 A FS) to 4.88 µA (70 mA FS) |
Current Accuracy: | ± 1% of full scale current of selected range |
Voltage Measurement and Data Acquisition: | |
Set and Read Voltage vs. WE: | ± 5.000 V |
Cell Voltage Sense Leads: | Differential with driven shields |
Voltage Measurement Resolution: | 152 µV |
Sense Lead Input Resistance: | 1 GΩ |
Data Acquisition Rate: | 10 points/second |
Impedance Analyzer: | |
Internal Impedance Analyzer Type: | Single sine, one generator and two gain/phase measurement channels |
Internal Analyzer Frequency Range: | 1 mHz to 100 kHz |
Electrolyte Fluid Handling System: | Computer-controlled peristaltic pumps (2); <1 to 1,000 mL/min |
All non-metallic flow path and reservoirs | |
Independent drain & purge valves for positive & negative electrolytes | |
Nitrogen purge and blanket; Electrolyte level monitoring | |
Temperature Controllers: | |
Three: | Cell, anode electrolyte, cathode electrolyte |
Set & Report Accuracy: | ±0.25% of span, ±1 least significant digit |
Sensor Type: | Thermocouple, Type T |
Environment: | |
Operating Temperature: | 5 to 35 °C |
Power Source: | 110-120 V, 50-60 Hz, 10 A (Export model 220-240 V, 50-60 Hz, 5 A) |
Size and Weight: | Electronic Control Unit: 48 W x 13 H x 53 D cm3 (19 x 5.25 x 21 in3); 9 kg (20 lb) |
Fluid Control Unit: | 50 W x 65 H x 65 D cm3 (20 x 25 x 25 in3); 16 kg (35 lb) |
Safety Features: | Automatic shutdown on under- and over-voltage or current, over-temperature, or communications failure |
Emergency stop switch for manual operator shutdown | |
*Specifications for 25o C ambient temperature unless otherwise noted. |